Description
Measuring range
|
0-100%
|
Output signal in air
|
13-16mV
|
Circuit interface
|
3Pin (Molex)
|
Accuracy and repeatability
|
<1% volume oxygen, after calibration in 100% oxygen
|
Linearity error
|
<3% relative error
|
Response time
|
<12 seconds, relative to 90% relative value
|
Zero drift
|
<200uV placed in 100% nitrogen for 5 minutes
|
Cross-talk: <0.5% volume oxygen for
|
10% CO2 in balanced nitrogen
80% N2O in balanced nitrogen 7.5% halothane in balanced nitrogen 7.5% isoflurane in balanced nitrogen 7.5% enflurane in balanced nitrogen 9% enflurane in balanced nitrogen 20% desflurane in balanced nitrogen |
Influence of temperature
|
-0.03% relative temperature, 25 ℃
|
Effect of pressure
|
Proportional to oxygen partial pressure
|
Effect of mechanical shock
|
<1% relative error, after falling from a height of 1 meter
|
Working temperature
|
0 ℃ -50 ℃
|
Temperature compensation
|
Built-in thermistor compensation
|
Temperature effect
|
25 ℃ -40 ℃, 3% relative error
|
Compensation (steady state)
|
8% relative error between 0 ° C and 50 ° C
|
Working humidity
|
0-99% RH (for several hours in condensable air)
|